A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature

We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors...

Full description

Saved in:
Bibliographic Details
Main Authors: David Morley, Ian Starks, Justin Lim
Format: Article
Language:English
Published: Wiley 2012-01-01
Series:Case Reports in Orthopedics
Online Access:http://dx.doi.org/10.1155/2012/158604
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors and narrow tapers should be used with caution, especially in heavy, active patients.
ISSN:2090-6749