Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases
Introduction: The aim of this study was to evaluate and measure the effectiveness of a particular radiotherapy treatment planning system, namely 3D conformal radiation therapy (3D-CRT), and radiobiological effects in postmastectomy breast cancer axillary lymph metastases using a patient-specific 3D...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Mashhad University of Medical Sciences
2024-10-01
|
Series: | Iranian Journal of Medical Physics |
Subjects: | |
Online Access: | https://ijmp.mums.ac.ir/article_23834_919f6acdded9e3b7f65917a336291a35.pdf |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832595702332522496 |
---|---|
author | Endarko Endarko Aditya Hariyanto Siti Aisyah Nandia Kavilani Ahmad Syafi'i |
author_facet | Endarko Endarko Aditya Hariyanto Siti Aisyah Nandia Kavilani Ahmad Syafi'i |
author_sort | Endarko Endarko |
collection | DOAJ |
description | Introduction: The aim of this study was to evaluate and measure the effectiveness of a particular radiotherapy treatment planning system, namely 3D conformal radiation therapy (3D-CRT), and radiobiological effects in postmastectomy breast cancer axillary lymph metastases using a patient-specific 3D printed anthropomorphic phantom of the right breast after surgery.
Material and Methods: The chest wall and axilla were the two treatment targets for 3D-CRT planning techniques. Each target was subjected to variations of two and three irradiation fields, with a dose of 50.00 Gy and fractionation of 2.00 Gy. From the variations, the radiation dose received by planning target volume (PTV) chest wall, PTV-Axillary, and organs at risk (OARs) was evaluated to determine the best planning technique through dosimetry verification. The thermoluminescent dosimeter (TLD) 100 Chip and EBT3 film were employed to measure the radiation dose at the PTV and OARs.
Results: The planning system evaluation based on dose volume histogram showed no significant difference (p = 0.993) between two and three irradiation fields in both the chest wall and axillary planning 3D-CRT technique. Normal tissue complication probability for OARs has values below 5%, while the smallest value was obtained for the left lung. Finally, the point dose verification between planning and measurement using TLD100 and EBT3 film indicated an average difference of 9.14% and 4.34%, respectively, with no significant difference.
Conclusion: This study successfully demonstrated the dosimetry evaluation and radiobiological effects of postmastectomy right breast cancer treatment using a patient-specific 3D printed anthropomorphic phantom, which met the established standards. |
format | Article |
id | doaj-art-a8e9f12add56486191cbb4aa4e3d0710 |
institution | Kabale University |
issn | 2345-3672 |
language | English |
publishDate | 2024-10-01 |
publisher | Mashhad University of Medical Sciences |
record_format | Article |
series | Iranian Journal of Medical Physics |
spelling | doaj-art-a8e9f12add56486191cbb4aa4e3d07102025-01-18T07:29:18ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722024-10-0121531432210.22038/ijmp.2024.73930.230923834Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary MetastasesEndarko Endarko0Aditya Hariyanto1Siti Aisyah2Nandia Kavilani3Ahmad Syafi'i4Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS - Sukolilo Surabaya 60111, East Java, Indonesia.Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS - Sukolilo Surabaya 60111, East Java, Indonesia.Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS - Sukolilo Surabaya 60111, East Java, Indonesia.Medical Physicist of Radiotherapy Installation, Naval Hospital Dr. Ramelan, Surabaya 60244, East Java, IndonesiaMedical Physicist of Radiotherapy Installation, Naval Hospital Dr. Ramelan, Surabaya 60244, East Java, IndonesiaIntroduction: The aim of this study was to evaluate and measure the effectiveness of a particular radiotherapy treatment planning system, namely 3D conformal radiation therapy (3D-CRT), and radiobiological effects in postmastectomy breast cancer axillary lymph metastases using a patient-specific 3D printed anthropomorphic phantom of the right breast after surgery. Material and Methods: The chest wall and axilla were the two treatment targets for 3D-CRT planning techniques. Each target was subjected to variations of two and three irradiation fields, with a dose of 50.00 Gy and fractionation of 2.00 Gy. From the variations, the radiation dose received by planning target volume (PTV) chest wall, PTV-Axillary, and organs at risk (OARs) was evaluated to determine the best planning technique through dosimetry verification. The thermoluminescent dosimeter (TLD) 100 Chip and EBT3 film were employed to measure the radiation dose at the PTV and OARs. Results: The planning system evaluation based on dose volume histogram showed no significant difference (p = 0.993) between two and three irradiation fields in both the chest wall and axillary planning 3D-CRT technique. Normal tissue complication probability for OARs has values below 5%, while the smallest value was obtained for the left lung. Finally, the point dose verification between planning and measurement using TLD100 and EBT3 film indicated an average difference of 9.14% and 4.34%, respectively, with no significant difference. Conclusion: This study successfully demonstrated the dosimetry evaluation and radiobiological effects of postmastectomy right breast cancer treatment using a patient-specific 3D printed anthropomorphic phantom, which met the established standards.https://ijmp.mums.ac.ir/article_23834_919f6acdded9e3b7f65917a336291a35.pdfradiation therapy threedimensional conformal radiation therapy (3dcrt) phantom breast cancer dosimetry radiation protection |
spellingShingle | Endarko Endarko Aditya Hariyanto Siti Aisyah Nandia Kavilani Ahmad Syafi'i Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases Iranian Journal of Medical Physics radiation therapy three dimensional conformal radiation therapy (3d crt) phantom breast cancer dosimetry radiation protection |
title | Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases |
title_full | Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases |
title_fullStr | Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases |
title_full_unstemmed | Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases |
title_short | Evaluation of 3D-CRT Radiotherapy Treatment Planning with Radiobiological Effects in Patient-Specific 3D Printed Anthropomorphic Phantom Postmastectomy Breast Cancer Axillary Metastases |
title_sort | evaluation of 3d crt radiotherapy treatment planning with radiobiological effects in patient specific 3d printed anthropomorphic phantom postmastectomy breast cancer axillary metastases |
topic | radiation therapy three dimensional conformal radiation therapy (3d crt) phantom breast cancer dosimetry radiation protection |
url | https://ijmp.mums.ac.ir/article_23834_919f6acdded9e3b7f65917a336291a35.pdf |
work_keys_str_mv | AT endarkoendarko evaluationof3dcrtradiotherapytreatmentplanningwithradiobiologicaleffectsinpatientspecific3dprintedanthropomorphicphantompostmastectomybreastcanceraxillarymetastases AT adityahariyanto evaluationof3dcrtradiotherapytreatmentplanningwithradiobiologicaleffectsinpatientspecific3dprintedanthropomorphicphantompostmastectomybreastcanceraxillarymetastases AT sitiaisyah evaluationof3dcrtradiotherapytreatmentplanningwithradiobiologicaleffectsinpatientspecific3dprintedanthropomorphicphantompostmastectomybreastcanceraxillarymetastases AT nandiakavilani evaluationof3dcrtradiotherapytreatmentplanningwithradiobiologicaleffectsinpatientspecific3dprintedanthropomorphicphantompostmastectomybreastcanceraxillarymetastases AT ahmadsyafi039i evaluationof3dcrtradiotherapytreatmentplanningwithradiobiologicaleffectsinpatientspecific3dprintedanthropomorphicphantompostmastectomybreastcanceraxillarymetastases |