Metamaterial‐Enabled Hybrid Receive Coil for Enhanced Magnetic Resonance Imaging Capabilities

Abstract Magnetic resonance imaging (MRI) relies on high‐performance receive coils to achieve optimal signal‐to‐noise ratio (SNR), but conventional designs are often bulky and complex. Recent advancements in metamaterial technology have led to the development of metamaterial‐inspired receive coils t...