A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders

Introduction: Cortisol concentration is often estimated by competitive chemiluminescent immunoassays, which are prone to interference due to cross-reactivity. The extent of interference is inconsistent across different adrenal disorders and among different analyzers. Although liquid chromatography-m...

Full description

Saved in:
Bibliographic Details
Main Authors: Neeti Agrawal, Anirban Sinha, Partha P. Chakraborty, Rana Bhattacharjee, Avivar Awasthi, Animesh Maiti
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2025-01-01
Series:Indian Journal of Endocrinology and Metabolism
Subjects:
Online Access:https://journals.lww.com/10.4103/ijem.ijem_482_23
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850052915317702656
author Neeti Agrawal
Anirban Sinha
Partha P. Chakraborty
Rana Bhattacharjee
Avivar Awasthi
Animesh Maiti
author_facet Neeti Agrawal
Anirban Sinha
Partha P. Chakraborty
Rana Bhattacharjee
Avivar Awasthi
Animesh Maiti
author_sort Neeti Agrawal
collection DOAJ
description Introduction: Cortisol concentration is often estimated by competitive chemiluminescent immunoassays, which are prone to interference due to cross-reactivity. The extent of interference is inconsistent across different adrenal disorders and among different analyzers. Although liquid chromatography-mass spectrometry (LC-MS/MS) possesses better analytical specificity, it is not widely available. We aimed to compare cortisol values measured across four different analyzers across different adrenal disorders. Methods: Cortisol concentrations were measured in the serum of patients with congenital adrenal hyperplasia (CAH) (n = 12), primary adrenal insufficiency (PAI) (n = 11), endogenous Cushing’s syndrome (CS) (n = 6), adrenal incidentaloma (AI) (n = 8), and healthy subjects (HS) (n = 10) in the following analyzers: Advia Centaur XP (Siemens) (SACXP), Immulite1000 (Siemens) (SI1000), Cobas e411 (Roche) (RCe411), and Architect (Abbott)(AA). Results: In CAH patients, a poor agreement was observed between SAC XP [median (IQR) 14.6 (4.7) μg/dL] and RC e411 [median (IQR) 4.6 (3.9) μg/dL] [ICC: −0.016, (−0.55, 0.55)]. The correlation was also poor between SAC XP and SI 1000 [ICC: 0.00, (0.558, 0.551)] as well as between SAC XP and AA [ICC: 0.089, (−0.488, 0.612)]. The agreement was good between RCe411, SI 1000 [median (IQR) 9.6 (3) μg/dL], and AA [median (IQR) 5.1 (3.4) μg/dL] platforms in the same group. Measured cortisol values correlated well across all analyzers in PAI, CS, AI, and HS. Conclusions: Cortisol concentration demonstrated variable agreement among different analyzers in different adrenal disorders. In CAH, cortisol values measured by SAC XP poorly correlate with RC e411, SI 1000, and AA platforms. Reassessment in another analyzer therefore would be prudent to avoid devastating consequences of unrecognized hypocortisolism in this subset of patients. A good concordance was observed among platforms in other conditions.
format Article
id doaj-art-ffa6d6bb4c2c4345b037cb6a0d82c3aa
institution DOAJ
issn 2230-8210
2230-9500
language English
publishDate 2025-01-01
publisher Wolters Kluwer Medknow Publications
record_format Article
series Indian Journal of Endocrinology and Metabolism
spelling doaj-art-ffa6d6bb4c2c4345b037cb6a0d82c3aa2025-08-20T02:52:41ZengWolters Kluwer Medknow PublicationsIndian Journal of Endocrinology and Metabolism2230-82102230-95002025-01-01291838810.4103/ijem.ijem_482_23A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal DisordersNeeti AgrawalAnirban SinhaPartha P. ChakrabortyRana BhattacharjeeAvivar AwasthiAnimesh MaitiIntroduction: Cortisol concentration is often estimated by competitive chemiluminescent immunoassays, which are prone to interference due to cross-reactivity. The extent of interference is inconsistent across different adrenal disorders and among different analyzers. Although liquid chromatography-mass spectrometry (LC-MS/MS) possesses better analytical specificity, it is not widely available. We aimed to compare cortisol values measured across four different analyzers across different adrenal disorders. Methods: Cortisol concentrations were measured in the serum of patients with congenital adrenal hyperplasia (CAH) (n = 12), primary adrenal insufficiency (PAI) (n = 11), endogenous Cushing’s syndrome (CS) (n = 6), adrenal incidentaloma (AI) (n = 8), and healthy subjects (HS) (n = 10) in the following analyzers: Advia Centaur XP (Siemens) (SACXP), Immulite1000 (Siemens) (SI1000), Cobas e411 (Roche) (RCe411), and Architect (Abbott)(AA). Results: In CAH patients, a poor agreement was observed between SAC XP [median (IQR) 14.6 (4.7) μg/dL] and RC e411 [median (IQR) 4.6 (3.9) μg/dL] [ICC: −0.016, (−0.55, 0.55)]. The correlation was also poor between SAC XP and SI 1000 [ICC: 0.00, (0.558, 0.551)] as well as between SAC XP and AA [ICC: 0.089, (−0.488, 0.612)]. The agreement was good between RCe411, SI 1000 [median (IQR) 9.6 (3) μg/dL], and AA [median (IQR) 5.1 (3.4) μg/dL] platforms in the same group. Measured cortisol values correlated well across all analyzers in PAI, CS, AI, and HS. Conclusions: Cortisol concentration demonstrated variable agreement among different analyzers in different adrenal disorders. In CAH, cortisol values measured by SAC XP poorly correlate with RC e411, SI 1000, and AA platforms. Reassessment in another analyzer therefore would be prudent to avoid devastating consequences of unrecognized hypocortisolism in this subset of patients. A good concordance was observed among platforms in other conditions.https://journals.lww.com/10.4103/ijem.ijem_482_23analyzerchemiluminescent immunoassaycongenital adrenal hyperplasiacortisolliquid chromatography-mass spectrometry
spellingShingle Neeti Agrawal
Anirban Sinha
Partha P. Chakraborty
Rana Bhattacharjee
Avivar Awasthi
Animesh Maiti
A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders
Indian Journal of Endocrinology and Metabolism
analyzer
chemiluminescent immunoassay
congenital adrenal hyperplasia
cortisol
liquid chromatography-mass spectrometry
title A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders
title_full A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders
title_fullStr A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders
title_full_unstemmed A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders
title_short A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders
title_sort cross sectional study on comparison of serum cortisol concentration measured by chemiluminescent immunoassay in four different automated analyzers in a variety of adrenal disorders
topic analyzer
chemiluminescent immunoassay
congenital adrenal hyperplasia
cortisol
liquid chromatography-mass spectrometry
url https://journals.lww.com/10.4103/ijem.ijem_482_23
work_keys_str_mv AT neetiagrawal acrosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT anirbansinha acrosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT parthapchakraborty acrosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT ranabhattacharjee acrosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT avivarawasthi acrosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT animeshmaiti acrosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT neetiagrawal crosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT anirbansinha crosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT parthapchakraborty crosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT ranabhattacharjee crosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT avivarawasthi crosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders
AT animeshmaiti crosssectionalstudyoncomparisonofserumcortisolconcentrationmeasuredbychemiluminescentimmunoassayinfourdifferentautomatedanalyzersinavarietyofadrenaldisorders