Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes
Pregnant ewes are susceptible to hypoglycemia and ketosis; therefore, monitoring glycemic status is extremely important. Portable blood glucose meters (PBGMs) can assist in quickly and conveniently identifying glycemic disturbances in this species, provided that they meet the criteria of analytical...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2025-01-01
|
Series: | Veterinary Sciences |
Subjects: | |
Online Access: | https://www.mdpi.com/2306-7381/12/1/47 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832587402041884672 |
---|---|
author | José Lucas Xavier Lopes Raquel Fraga e Silva Raimondo Luiza Rodegheri Jacondino Beatriz Riet Correa Clara Satsuki Mori Álan Gomes Pöppl |
author_facet | José Lucas Xavier Lopes Raquel Fraga e Silva Raimondo Luiza Rodegheri Jacondino Beatriz Riet Correa Clara Satsuki Mori Álan Gomes Pöppl |
author_sort | José Lucas Xavier Lopes |
collection | DOAJ |
description | Pregnant ewes are susceptible to hypoglycemia and ketosis; therefore, monitoring glycemic status is extremely important. Portable blood glucose meters (PBGMs) can assist in quickly and conveniently identifying glycemic disturbances in this species, provided that they meet the criteria of analytical accuracy. This study evaluated the performance of a human PBGM (Accu-Chek Performa<sup>®</sup>, Roche Diagnostics, Basel, Switzerland) in the glycemic evaluation of 34 pregnant ewes at days 90 and 120 of pregnancy in comparison with the results of glycemia determination by a reference method (RM). The device showed a high positive correlation (r = 0.71, 95%CI = 0.57–0.82, <i>p</i> < 0.0001) with the RM; however, 96.6% of the PBGM results (58.5 ± 9.82 mg/dL) were higher (<i>p</i> < 0.0001) than those obtained in the laboratory (48.6 ± 9.31 mg/dL). The PBGM tested was considered analytically inaccurate according to ISO 15197:2013, which states that when glucose levels are below 100 mg/dL, 95% of the measurements should deviate by no more than 15 mg/dL from the RM value, and 1/3 of the PBGM results were above this limit. Hypoglycemia (<50 mg/dL) was documented in 60.29% of samples tested on with the RM, but only 17.64% of results were below 50 mg/dL using the PBGM. Due to these limitations, Accu-Check Performa<sup>®</sup> results should be interpreted cautiously in pregnant sheep suspected of hypoglycemia. |
format | Article |
id | doaj-art-ff03a77bae674ba7a533a64c36a6e37d |
institution | Kabale University |
issn | 2306-7381 |
language | English |
publishDate | 2025-01-01 |
publisher | MDPI AG |
record_format | Article |
series | Veterinary Sciences |
spelling | doaj-art-ff03a77bae674ba7a533a64c36a6e37d2025-01-24T13:52:06ZengMDPI AGVeterinary Sciences2306-73812025-01-011214710.3390/vetsci12010047Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant EwesJosé Lucas Xavier Lopes0Raquel Fraga e Silva Raimondo1Luiza Rodegheri Jacondino2Beatriz Riet Correa3Clara Satsuki Mori4Álan Gomes Pöppl5Programa de Pós-Graduação em Ciências Veterinárias, Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre 91540-000, BrazilNúcleo RuminAção, Ensino, Pesquisa e Extensão em Ruminantes, Faculdade de Veterinária, Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre 91540-000, BrazilPrograma de Pós-Graduação em Zootecnia, Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre 91540-000, BrazilNúcleo RuminAção, Ensino, Pesquisa e Extensão em Ruminantes, Faculdade de Veterinária, Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre 91540-000, BrazilPrograma de Pós-Graduação em Clínica Veterinária, Faculdade de Medicina Veterinária e Zootecnia, Universidade de São Paulo (USP), São Paulo 05508-270, BrazilPrograma de Pós-Graduação em Ciências Veterinárias, Universidade Federal do Rio Grande do Sul (UFRGS), Porto Alegre 91540-000, BrazilPregnant ewes are susceptible to hypoglycemia and ketosis; therefore, monitoring glycemic status is extremely important. Portable blood glucose meters (PBGMs) can assist in quickly and conveniently identifying glycemic disturbances in this species, provided that they meet the criteria of analytical accuracy. This study evaluated the performance of a human PBGM (Accu-Chek Performa<sup>®</sup>, Roche Diagnostics, Basel, Switzerland) in the glycemic evaluation of 34 pregnant ewes at days 90 and 120 of pregnancy in comparison with the results of glycemia determination by a reference method (RM). The device showed a high positive correlation (r = 0.71, 95%CI = 0.57–0.82, <i>p</i> < 0.0001) with the RM; however, 96.6% of the PBGM results (58.5 ± 9.82 mg/dL) were higher (<i>p</i> < 0.0001) than those obtained in the laboratory (48.6 ± 9.31 mg/dL). The PBGM tested was considered analytically inaccurate according to ISO 15197:2013, which states that when glucose levels are below 100 mg/dL, 95% of the measurements should deviate by no more than 15 mg/dL from the RM value, and 1/3 of the PBGM results were above this limit. Hypoglycemia (<50 mg/dL) was documented in 60.29% of samples tested on with the RM, but only 17.64% of results were below 50 mg/dL using the PBGM. Due to these limitations, Accu-Check Performa<sup>®</sup> results should be interpreted cautiously in pregnant sheep suspected of hypoglycemia.https://www.mdpi.com/2306-7381/12/1/47sheepblood glucoseAccu-Check PerformaISO 15197:2013 |
spellingShingle | José Lucas Xavier Lopes Raquel Fraga e Silva Raimondo Luiza Rodegheri Jacondino Beatriz Riet Correa Clara Satsuki Mori Álan Gomes Pöppl Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes Veterinary Sciences sheep blood glucose Accu-Check Performa ISO 15197:2013 |
title | Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes |
title_full | Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes |
title_fullStr | Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes |
title_full_unstemmed | Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes |
title_short | Diagnostic Limitations of Applying a Human Portable Blood Glucose Meter in the Detection of Hypoglycemia in Pregnant Ewes |
title_sort | diagnostic limitations of applying a human portable blood glucose meter in the detection of hypoglycemia in pregnant ewes |
topic | sheep blood glucose Accu-Check Performa ISO 15197:2013 |
url | https://www.mdpi.com/2306-7381/12/1/47 |
work_keys_str_mv | AT joselucasxavierlopes diagnosticlimitationsofapplyingahumanportablebloodglucosemeterinthedetectionofhypoglycemiainpregnantewes AT raquelfragaesilvaraimondo diagnosticlimitationsofapplyingahumanportablebloodglucosemeterinthedetectionofhypoglycemiainpregnantewes AT luizarodegherijacondino diagnosticlimitationsofapplyingahumanportablebloodglucosemeterinthedetectionofhypoglycemiainpregnantewes AT beatrizrietcorrea diagnosticlimitationsofapplyingahumanportablebloodglucosemeterinthedetectionofhypoglycemiainpregnantewes AT clarasatsukimori diagnosticlimitationsofapplyingahumanportablebloodglucosemeterinthedetectionofhypoglycemiainpregnantewes AT alangomespoppl diagnosticlimitationsofapplyingahumanportablebloodglucosemeterinthedetectionofhypoglycemiainpregnantewes |