Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT Systems

Objectives. SPECT/slow-rotation low-output CT systems can produce streak artifacts in filtered backprojection (FBP) attenuation maps, impacting attenuation correction (AC) in myocardial perfusion imaging. This paper presents an adaptive Bayesian iterative transmission reconstruction (ABITR) algorith...

Full description

Saved in:
Bibliographic Details
Main Authors: Ji Chen, Ernest V. Garcia, Russell D. Folks, Aharon Peretz, James R. Galt
Format: Article
Language:English
Published: Wiley 2007-01-01
Series:International Journal of Biomedical Imaging
Online Access:http://dx.doi.org/10.1155/2007/18709
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1832563122937790464
author Ji Chen
Ernest V. Garcia
Russell D. Folks
Aharon Peretz
James R. Galt
author_facet Ji Chen
Ernest V. Garcia
Russell D. Folks
Aharon Peretz
James R. Galt
author_sort Ji Chen
collection DOAJ
description Objectives. SPECT/slow-rotation low-output CT systems can produce streak artifacts in filtered backprojection (FBP) attenuation maps, impacting attenuation correction (AC) in myocardial perfusion imaging. This paper presents an adaptive Bayesian iterative transmission reconstruction (ABITR) algorithm for more accurate AC. Methods. In each iteration, ABITR calculated a three-dimensional prior containing the pixels with attenuation coefficients similar to water, then used it to encourage these pixels to the water value. ABITR was tested with a cardiac phantom and 4 normal patients acquired by a GE Millennium VG/Hawkeye system. Results. FBP AC and ABITR AC produced similar phantom results. For the patients, streak artifacts were observed in the FBP and ordered-subsets expectation-maximization (OSEM) maps but not in the ABITR maps, and ABITR AC produced more uniform images than FBP AC and OSEM AC. Conclusion. ABITR can improve the quality of the attenuation map, producing more uniform images for normal studies.
format Article
id doaj-art-ebc0077579f84470ac5cb2c013a5a700
institution Kabale University
issn 1687-4188
1687-4196
language English
publishDate 2007-01-01
publisher Wiley
record_format Article
series International Journal of Biomedical Imaging
spelling doaj-art-ebc0077579f84470ac5cb2c013a5a7002025-02-03T01:21:00ZengWileyInternational Journal of Biomedical Imaging1687-41881687-41962007-01-01200710.1155/2007/1870918709Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT SystemsJi Chen0Ernest V. Garcia1Russell D. Folks2Aharon Peretz3James R. Galt4Department of Radiology, School of Medicine, Emory University, 1364 Clifton Road, Atlanta 30322, GA, USADepartment of Radiology, School of Medicine, Emory University, 1364 Clifton Road, Atlanta 30322, GA, USADepartment of Radiology, School of Medicine, Emory University, 1364 Clifton Road, Atlanta 30322, GA, USAGE Healthcare, Haifa 39120, IsraelDepartment of Radiology, School of Medicine, Emory University, 1364 Clifton Road, Atlanta 30322, GA, USAObjectives. SPECT/slow-rotation low-output CT systems can produce streak artifacts in filtered backprojection (FBP) attenuation maps, impacting attenuation correction (AC) in myocardial perfusion imaging. This paper presents an adaptive Bayesian iterative transmission reconstruction (ABITR) algorithm for more accurate AC. Methods. In each iteration, ABITR calculated a three-dimensional prior containing the pixels with attenuation coefficients similar to water, then used it to encourage these pixels to the water value. ABITR was tested with a cardiac phantom and 4 normal patients acquired by a GE Millennium VG/Hawkeye system. Results. FBP AC and ABITR AC produced similar phantom results. For the patients, streak artifacts were observed in the FBP and ordered-subsets expectation-maximization (OSEM) maps but not in the ABITR maps, and ABITR AC produced more uniform images than FBP AC and OSEM AC. Conclusion. ABITR can improve the quality of the attenuation map, producing more uniform images for normal studies.http://dx.doi.org/10.1155/2007/18709
spellingShingle Ji Chen
Ernest V. Garcia
Russell D. Folks
Aharon Peretz
James R. Galt
Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT Systems
International Journal of Biomedical Imaging
title Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT Systems
title_full Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT Systems
title_fullStr Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT Systems
title_full_unstemmed Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT Systems
title_short Adaptive Bayesian Iterative Transmission Reconstruction for Attenuation Correction in Myocardial Perfusion Imaging with SPECT/Slow-Rotation Low-Output CT Systems
title_sort adaptive bayesian iterative transmission reconstruction for attenuation correction in myocardial perfusion imaging with spect slow rotation low output ct systems
url http://dx.doi.org/10.1155/2007/18709
work_keys_str_mv AT jichen adaptivebayesianiterativetransmissionreconstructionforattenuationcorrectioninmyocardialperfusionimagingwithspectslowrotationlowoutputctsystems
AT ernestvgarcia adaptivebayesianiterativetransmissionreconstructionforattenuationcorrectioninmyocardialperfusionimagingwithspectslowrotationlowoutputctsystems
AT russelldfolks adaptivebayesianiterativetransmissionreconstructionforattenuationcorrectioninmyocardialperfusionimagingwithspectslowrotationlowoutputctsystems
AT aharonperetz adaptivebayesianiterativetransmissionreconstructionforattenuationcorrectioninmyocardialperfusionimagingwithspectslowrotationlowoutputctsystems
AT jamesrgalt adaptivebayesianiterativetransmissionreconstructionforattenuationcorrectioninmyocardialperfusionimagingwithspectslowrotationlowoutputctsystems