MEASURING OF X-RAY SOURCE SIZE BY USING COMPOUND REFRACTIVE X-RAY LENS
Compound refractive lens was used for measuring size of 2-3 beamline Standford synchrotron radiation source and a size of microfocus X-ray tube. X-ray beam from the source was focused by the lens and parameters of the beam at image plane were measured. Scanning diaphragm and X-ray CCD-camera were us...
Saved in:
Main Author: | |
---|---|
Format: | Article |
Language: | English |
Published: |
Belarusian National Technical University
2015-04-01
|
Series: | Приборы и методы измерений |
Subjects: | |
Online Access: | https://pimi.bntu.by/jour/article/view/142 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Compound refractive lens was used for measuring size of 2-3 beamline Standford synchrotron radiation source and a size of microfocus X-ray tube. X-ray beam from the source was focused by the lens and parameters of the beam at image plane were measured. Scanning diaphragm and X-ray CCD-camera were used for measuring X-ray beam. It was found that the vertical size of synchrotron source is equal to 0,6 mm and the size of the X-ray tube focal spot is equal to 60 micrometers. |
---|---|
ISSN: | 2220-9506 2414-0473 |