Analysis of lenses absorbed dose in head CT scan with the use of bismuth shielding

Computed Tomography (CT) has become an important tool to diagnose cancer and to obtain additional information for different clinical questions. However, CT scan usually requires a higher radiation exposure than a conventional radiography examination. Head CT scans are used for diagnosis of traumatic...

Full description

Saved in:
Bibliographic Details
Main Authors: Fernanda Stephanie Santos, Priscila do Carmo Santana, Arnaldo Prata Mourão
Format: Article
Language:English
Published: Brazilian Radiation Protection Society (Sociedade Brasileira de Proteção Radiológica, SBPR) 2018-07-01
Series:Brazilian Journal of Radiation Sciences
Subjects:
Online Access:https://bjrs.org.br/revista/index.php/REVISTA/article/view/446
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Computed Tomography (CT) has become an important tool to diagnose cancer and to obtain additional information for different clinical questions. However, CT scan usually requires a higher radiation exposure than a conventional radiography examination. Head CT scans are used for diagnosis of traumatic head injuries, infections and other diseases with instability. Based on this information, it was studied the dose variation deposited in the lenses and in nearby organs, such as: pharynx, hypophysis and salivary gland with and without the use of bismuth shield. In this study a head CT scan was performed on anthropomorphic male phantom using a GE scanner. Dose measurements have been performed by using radiochromic film strips to register the individual doses in the organs of interest. The results show us that the lenses had a reduction of 26% of the dose with the use of the bismuth shield.
ISSN:2319-0612