Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site
Background/Purpose. We aimed to investigate the influence of the sampling site on the variability of ACT measurement. Activated clotting time (ACT) has been used for decades in cardiac surgery and interventional cardiology to assess unfractionated heparin activity. However, standardized protocols fo...
Saved in:
Main Authors: | , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2021-01-01
|
Series: | Journal of Interventional Cardiology |
Online Access: | http://dx.doi.org/10.1155/2021/4091289 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832566301218832384 |
---|---|
author | J. C Heemelaar T. Berkhout A. A. C. M. Heestermans J. C. Zant A. M. J. de Vos N. J. W. Verouden M. T. Dirksen J. van Ramshorst |
author_facet | J. C Heemelaar T. Berkhout A. A. C. M. Heestermans J. C. Zant A. M. J. de Vos N. J. W. Verouden M. T. Dirksen J. van Ramshorst |
author_sort | J. C Heemelaar |
collection | DOAJ |
description | Background/Purpose. We aimed to investigate the influence of the sampling site on the variability of ACT measurement. Activated clotting time (ACT) has been used for decades in cardiac surgery and interventional cardiology to assess unfractionated heparin activity. However, standardized protocols for the use of ACT measurement in the catheterization laboratory are lacking. Methods/Materials. After elective cardiac catheterization, ACT measurements were collected in simultaneously obtained blood samples from three different sample sites: the arterial catheter, arterial sheath, and peripheral intravenous line. Measurements were performed using the i-Stat® device (Abbott, Princeton, NJ, USA). The study was conducted with approval of the local medical ethical committee. Results. In 100 patients (mean age 67.1, 65% male), no significant differences were observed in ACT values obtained from the guiding catheter and arterial sheath (mean difference (MD) −18.3 s; standard deviation (SD) 96 s; P=0.067). Contrarily, ACT values obtained from the intravenous line were significantly lower as compared to values obtained from the guiding catheter (MD 25.7 s; SD 75.5; P=0.003) and arterial sheath (MD 39 s; SD 102.8; P<0.001). Furthermore, ACT measurements from the arterial sheath showed a statistically significant proportional bias when compared to the other sampling sites (sheath vs. catheter, r = 0.761, P=0.001; sheath vs. IVL, r = 1.013, P<0.001). Conclusions. The present study shows statistical significance and possibly clinically relevant variations between ACT measurements from different sample sites. Bias in ACT measurements may be minimized by using uniform protocols for ACT measurement during cardiac catheterization. |
format | Article |
id | doaj-art-ca59be4459bf45b89c9b7f036fd9a7be |
institution | Kabale University |
issn | 0896-4327 1540-8183 |
language | English |
publishDate | 2021-01-01 |
publisher | Wiley |
record_format | Article |
series | Journal of Interventional Cardiology |
spelling | doaj-art-ca59be4459bf45b89c9b7f036fd9a7be2025-02-03T01:04:33ZengWileyJournal of Interventional Cardiology0896-43271540-81832021-01-01202110.1155/2021/40912894091289Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling SiteJ. C Heemelaar0T. Berkhout1A. A. C. M. Heestermans2J. C. Zant3A. M. J. de Vos4N. J. W. Verouden5M. T. Dirksen6J. van Ramshorst7Department of Cardiology, Northwest Clinics, Alkmaar, NetherlandsDepartment of Cardiology, Northwest Clinics, Alkmaar, NetherlandsDepartment of Cardiology, Northwest Clinics, Alkmaar, NetherlandsDepartment of Clinical Chemistry, Northwest Clinics, Alkmaar, NetherlandsDepartment of Cardiology, Catharina Hospital, Eindhoven, NetherlandsDepartment of Cardiology, Amsterdam University Medical Center, Amsterdam, NetherlandsDepartment of Cardiology, Northwest Clinics, Alkmaar, NetherlandsDepartment of Cardiology, Northwest Clinics, Alkmaar, NetherlandsBackground/Purpose. We aimed to investigate the influence of the sampling site on the variability of ACT measurement. Activated clotting time (ACT) has been used for decades in cardiac surgery and interventional cardiology to assess unfractionated heparin activity. However, standardized protocols for the use of ACT measurement in the catheterization laboratory are lacking. Methods/Materials. After elective cardiac catheterization, ACT measurements were collected in simultaneously obtained blood samples from three different sample sites: the arterial catheter, arterial sheath, and peripheral intravenous line. Measurements were performed using the i-Stat® device (Abbott, Princeton, NJ, USA). The study was conducted with approval of the local medical ethical committee. Results. In 100 patients (mean age 67.1, 65% male), no significant differences were observed in ACT values obtained from the guiding catheter and arterial sheath (mean difference (MD) −18.3 s; standard deviation (SD) 96 s; P=0.067). Contrarily, ACT values obtained from the intravenous line were significantly lower as compared to values obtained from the guiding catheter (MD 25.7 s; SD 75.5; P=0.003) and arterial sheath (MD 39 s; SD 102.8; P<0.001). Furthermore, ACT measurements from the arterial sheath showed a statistically significant proportional bias when compared to the other sampling sites (sheath vs. catheter, r = 0.761, P=0.001; sheath vs. IVL, r = 1.013, P<0.001). Conclusions. The present study shows statistical significance and possibly clinically relevant variations between ACT measurements from different sample sites. Bias in ACT measurements may be minimized by using uniform protocols for ACT measurement during cardiac catheterization.http://dx.doi.org/10.1155/2021/4091289 |
spellingShingle | J. C Heemelaar T. Berkhout A. A. C. M. Heestermans J. C. Zant A. M. J. de Vos N. J. W. Verouden M. T. Dirksen J. van Ramshorst Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site Journal of Interventional Cardiology |
title | Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site |
title_full | Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site |
title_fullStr | Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site |
title_full_unstemmed | Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site |
title_short | Measuring Activated Clotting Time during Cardiac Catheterization and PCI: The Effect of the Sampling Site |
title_sort | measuring activated clotting time during cardiac catheterization and pci the effect of the sampling site |
url | http://dx.doi.org/10.1155/2021/4091289 |
work_keys_str_mv | AT jcheemelaar measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite AT tberkhout measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite AT aacmheestermans measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite AT jczant measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite AT amjdevos measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite AT njwverouden measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite AT mtdirksen measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite AT jvanramshorst measuringactivatedclottingtimeduringcardiaccatheterizationandpcitheeffectofthesamplingsite |