A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature

We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors...

Full description

Saved in:
Bibliographic Details
Main Authors: David Morley, Ian Starks, Justin Lim
Format: Article
Language:English
Published: Wiley 2012-01-01
Series:Case Reports in Orthopedics
Online Access:http://dx.doi.org/10.1155/2012/158604
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1832551426102919168
author David Morley
Ian Starks
Justin Lim
author_facet David Morley
Ian Starks
Justin Lim
author_sort David Morley
collection DOAJ
description We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors and narrow tapers should be used with caution, especially in heavy, active patients.
format Article
id doaj-art-c973923e45004a30bccb769656e86739
institution Kabale University
issn 2090-6749
2090-6757
language English
publishDate 2012-01-01
publisher Wiley
record_format Article
series Case Reports in Orthopedics
spelling doaj-art-c973923e45004a30bccb769656e867392025-02-03T06:01:30ZengWileyCase Reports in Orthopedics2090-67492090-67572012-01-01201210.1155/2012/158604158604A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the LiteratureDavid Morley0Ian Starks1Justin Lim2City General Hospital, University Hospital of North Staffordshire, Newcastle Road, Stoke-on-Trent ST4 6QG, UKCity General Hospital, University Hospital of North Staffordshire, Newcastle Road, Stoke-on-Trent ST4 6QG, UKCity General Hospital, University Hospital of North Staffordshire, Newcastle Road, Stoke-on-Trent ST4 6QG, UKWe report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors and narrow tapers should be used with caution, especially in heavy, active patients.http://dx.doi.org/10.1155/2012/158604
spellingShingle David Morley
Ian Starks
Justin Lim
A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
Case Reports in Orthopedics
title A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
title_full A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
title_fullStr A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
title_full_unstemmed A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
title_short A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
title_sort case of a c stem fracture at the head neck junction and a review of the literature
url http://dx.doi.org/10.1155/2012/158604
work_keys_str_mv AT davidmorley acaseofacstemfractureattheheadneckjunctionandareviewoftheliterature
AT ianstarks acaseofacstemfractureattheheadneckjunctionandareviewoftheliterature
AT justinlim acaseofacstemfractureattheheadneckjunctionandareviewoftheliterature
AT davidmorley caseofacstemfractureattheheadneckjunctionandareviewoftheliterature
AT ianstarks caseofacstemfractureattheheadneckjunctionandareviewoftheliterature
AT justinlim caseofacstemfractureattheheadneckjunctionandareviewoftheliterature