Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device

There is an increasing prevalence of patients with concomitant implantable cardioverter-defibrillators (ICDs) and left ventricular devices (LVADs). The potential for negative interactions between these continually evolving technologies is a valid concern. Previously reported interactions include ina...

Full description

Saved in:
Bibliographic Details
Main Authors: Sebhat Erqou, Robert L. Kormos, Norman C. Wang, Dennis M. McNamara, Raveen Bazaz
Format: Article
Language:English
Published: Wiley 2018-01-01
Series:Case Reports in Cardiology
Online Access:http://dx.doi.org/10.1155/2018/6195045
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:There is an increasing prevalence of patients with concomitant implantable cardioverter-defibrillators (ICDs) and left ventricular devices (LVADs). The potential for negative interactions between these continually evolving technologies is a valid concern. Previously reported interactions include inappropriate ICD therapy and interference with ICD telemetry function. Understanding the nature of such interactions and developing a comprehensive strategy to approach such situations are important. In this report, we describe a case of electromagnetic interference from LVAD inhibiting the pacing function of an ICD that was corrected by reprograming the device. We would encourage investigators to review patients with ICD and LVAD in their institutions in order to help assess the frequency and nature of these and other interactions.
ISSN:2090-6404
2090-6412