Auxetic Meta-Biomaterials: Computer Simulation and Experimental Results
One of the issues the modern hip implants face is that one side of the implant may detach due to stretching during its use. This leads to implant transverse compression and separation from the bone. This issue can be solved by using complex implants having one of the sides made of auxetic meta-bioma...
Saved in:
| Main Authors: | , , |
|---|---|
| Format: | Article |
| Language: | English |
| Published: |
MDPI AG
2025-02-01
|
| Series: | Metals |
| Subjects: | |
| Online Access: | https://www.mdpi.com/2075-4701/15/3/241 |
| Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
| Summary: | One of the issues the modern hip implants face is that one side of the implant may detach due to stretching during its use. This leads to implant transverse compression and separation from the bone. This issue can be solved by using complex implants having one of the sides made of auxetic meta-biomaterials with negative Poisson’s ratio. On the contrary, the cross-section of such materials being stretched will increase, which results in bone growth stimulation and minimum possibility of implant detachment. The aim of this paper is to design and fabricate titanium alloy auxetic meta-biomaterials based on 3D unit cells with three types of topologies. The works involved computer simulation to determine the expected properties of the samples. The samples were fabricated by the selective laser melting method and their properties were determined. Auxetic meta-biomaterials with Poisson’s ratio values of −0.09 and −0.003 and elastic modulus values typical for a human trabecular bone were fabricated in the course of the works. |
|---|---|