Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT
Objective. To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT). Methods. In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectra...
Saved in:
Main Authors: | , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2024-01-01
|
Series: | Journal of Ophthalmology |
Online Access: | http://dx.doi.org/10.1155/2024/3684626 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832559716083957760 |
---|---|
author | Hao Wu Zuocheng Wang Pengfei Wang Yifei Meng Zengying Wang Yuhong Xue Bohua Jiang Shuaixi Pan Zhipeng Yan |
author_facet | Hao Wu Zuocheng Wang Pengfei Wang Yifei Meng Zengying Wang Yuhong Xue Bohua Jiang Shuaixi Pan Zhipeng Yan |
author_sort | Hao Wu |
collection | DOAJ |
description | Objective. To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT). Methods. In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired t-test and Bland–Altman plots were used to analyze the differences and agreements of central vault measured by two devices. Results. Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired t-test showed that central vault measured by Tomey OA-2000 biometer was −7.25 ± 23.57 microns lower than that measured by Spectralis OCT (P=0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was −38.94 to 53.44 μm. Conclusion. Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements. |
format | Article |
id | doaj-art-ae832b7e9a10402d98a86d5d6f030680 |
institution | Kabale University |
issn | 2090-0058 |
language | English |
publishDate | 2024-01-01 |
publisher | Wiley |
record_format | Article |
series | Journal of Ophthalmology |
spelling | doaj-art-ae832b7e9a10402d98a86d5d6f0306802025-02-03T01:29:29ZengWileyJournal of Ophthalmology2090-00582024-01-01202410.1155/2024/3684626Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCTHao Wu0Zuocheng Wang1Pengfei Wang2Yifei Meng3Zengying Wang4Yuhong Xue5Bohua Jiang6Shuaixi Pan7Zhipeng Yan8Department of OphthalmologySchool of ManagementDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyObjective. To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT). Methods. In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired t-test and Bland–Altman plots were used to analyze the differences and agreements of central vault measured by two devices. Results. Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired t-test showed that central vault measured by Tomey OA-2000 biometer was −7.25 ± 23.57 microns lower than that measured by Spectralis OCT (P=0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was −38.94 to 53.44 μm. Conclusion. Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements.http://dx.doi.org/10.1155/2024/3684626 |
spellingShingle | Hao Wu Zuocheng Wang Pengfei Wang Yifei Meng Zengying Wang Yuhong Xue Bohua Jiang Shuaixi Pan Zhipeng Yan Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT Journal of Ophthalmology |
title | Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT |
title_full | Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT |
title_fullStr | Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT |
title_full_unstemmed | Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT |
title_short | Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT |
title_sort | repeatability and agreement of central vault for implantable collamer lens obtained by the tomey oa 2000 biometer and spectralis oct |
url | http://dx.doi.org/10.1155/2024/3684626 |
work_keys_str_mv | AT haowu repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT zuochengwang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT pengfeiwang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT yifeimeng repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT zengyingwang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT yuhongxue repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT bohuajiang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT shuaixipan repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct AT zhipengyan repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct |