Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT

Objective. To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT). Methods. In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectra...

Full description

Saved in:
Bibliographic Details
Main Authors: Hao Wu, Zuocheng Wang, Pengfei Wang, Yifei Meng, Zengying Wang, Yuhong Xue, Bohua Jiang, Shuaixi Pan, Zhipeng Yan
Format: Article
Language:English
Published: Wiley 2024-01-01
Series:Journal of Ophthalmology
Online Access:http://dx.doi.org/10.1155/2024/3684626
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1832559716083957760
author Hao Wu
Zuocheng Wang
Pengfei Wang
Yifei Meng
Zengying Wang
Yuhong Xue
Bohua Jiang
Shuaixi Pan
Zhipeng Yan
author_facet Hao Wu
Zuocheng Wang
Pengfei Wang
Yifei Meng
Zengying Wang
Yuhong Xue
Bohua Jiang
Shuaixi Pan
Zhipeng Yan
author_sort Hao Wu
collection DOAJ
description Objective. To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT). Methods. In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired t-test and Bland–Altman plots were used to analyze the differences and agreements of central vault measured by two devices. Results. Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired t-test showed that central vault measured by Tomey OA-2000 biometer was −7.25 ± 23.57 microns lower than that measured by Spectralis OCT (P=0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was −38.94 to 53.44 μm. Conclusion. Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements.
format Article
id doaj-art-ae832b7e9a10402d98a86d5d6f030680
institution Kabale University
issn 2090-0058
language English
publishDate 2024-01-01
publisher Wiley
record_format Article
series Journal of Ophthalmology
spelling doaj-art-ae832b7e9a10402d98a86d5d6f0306802025-02-03T01:29:29ZengWileyJournal of Ophthalmology2090-00582024-01-01202410.1155/2024/3684626Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCTHao Wu0Zuocheng Wang1Pengfei Wang2Yifei Meng3Zengying Wang4Yuhong Xue5Bohua Jiang6Shuaixi Pan7Zhipeng Yan8Department of OphthalmologySchool of ManagementDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyDepartment of OphthalmologyObjective. To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT). Methods. In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired t-test and Bland–Altman plots were used to analyze the differences and agreements of central vault measured by two devices. Results. Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired t-test showed that central vault measured by Tomey OA-2000 biometer was −7.25 ± 23.57 microns lower than that measured by Spectralis OCT (P=0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was −38.94 to 53.44 μm. Conclusion. Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements.http://dx.doi.org/10.1155/2024/3684626
spellingShingle Hao Wu
Zuocheng Wang
Pengfei Wang
Yifei Meng
Zengying Wang
Yuhong Xue
Bohua Jiang
Shuaixi Pan
Zhipeng Yan
Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT
Journal of Ophthalmology
title Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT
title_full Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT
title_fullStr Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT
title_full_unstemmed Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT
title_short Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT
title_sort repeatability and agreement of central vault for implantable collamer lens obtained by the tomey oa 2000 biometer and spectralis oct
url http://dx.doi.org/10.1155/2024/3684626
work_keys_str_mv AT haowu repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT zuochengwang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT pengfeiwang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT yifeimeng repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT zengyingwang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT yuhongxue repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT bohuajiang repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT shuaixipan repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct
AT zhipengyan repeatabilityandagreementofcentralvaultforimplantablecollamerlensobtainedbythetomeyoa2000biometerandspectralisoct