Over-Response Correction of SNC 350p ROOS and PTW Markus Chambers in the Surface Dose and Build-Up Region Dosimetry of 6MV Photon Beam

Introduction: The design characteristics of ionization chambers are critical for the accurate measurement of Percentage Surface Dose (PSD) and Percentage Depth Dose (PDD) in the build-up region. This study assesses the over-response correction in PSD and PDD for different field sizes, using Markus a...

Full description

Saved in:
Bibliographic Details
Main Authors: Deepali Patil, Mukesh Zope, DINESH SAROJ, Aslam P. A., Basil George
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2024-07-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:https://ijmp.mums.ac.ir/article_22854_146b5af4eb242f46f7bc0b3ee66ca986.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Introduction: The design characteristics of ionization chambers are critical for the accurate measurement of Percentage Surface Dose (PSD) and Percentage Depth Dose (PDD) in the build-up region. This study assesses the over-response correction in PSD and PDD for different field sizes, using Markus and ROOS fixed separation parallel-plate ionization chambers. Material and Methods: The Sun Nuclear Corporation SNC 350P ROOS and PTW Markus parallel plate ionization chambers were used to measure 6 MV photon beam PDD along the central axis in the build-up region.The experiment was performed on a Solid Water® HE phantom with a density of 1.032 g/cm3.The Measurements were taken at a Source Surface distance (SSD) of 100 cm with Field sizes of 10x10cm2,20x20 cm2, and 30x30cm2 and at variable depths. Results: For field sizes of 10x10 cm2, 20x20 cm2, and 30x30 cm2 at a source-to-axis distance (SAD) of 100 cm, the PSD values obtained with the Markus chamber, before and after applying Rawlinson's equation for over-response correction, were as follows: 35.11%, 42.92%, 50.72% (before correction) and 28.37%, 36.19%, 44.05% (after correction). Similarly, the PSD values measured using the ROOS chamber, before and after applying over-response correction, were as follows: 41.61%, 50.50%, 58.04% (before correction), and 33.96%, 42.87%, 50.38% (after correction). Conclusion: The application of over-response correction using Rawlinson's equation improved the accuracy of measurements for both chambers, emphasizing the significance of employing appropriate correction techniques to ensure precise dose measurements in the build-up region using ionization chambers.
ISSN:2345-3672