Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses

Background. The purpose of this observational case series study was to investigate the role of 25 MHz ultrasound biomicroscopy (UBM) in detecting subluxated lenses and compare it with 50 MHz UBM. Methods. 45 patients (49 eyes) with suspected subluxation of the lens and 20 normal volunteers (40 eyes)...

Full description

Saved in:
Bibliographic Details
Main Authors: Mingyu Shi, Liwei Ma, Jinsong Zhang, Qichang Yan
Format: Article
Language:English
Published: Wiley 2018-01-01
Series:Journal of Ophthalmology
Online Access:http://dx.doi.org/10.1155/2018/3760280
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1832557120160006144
author Mingyu Shi
Liwei Ma
Jinsong Zhang
Qichang Yan
author_facet Mingyu Shi
Liwei Ma
Jinsong Zhang
Qichang Yan
author_sort Mingyu Shi
collection DOAJ
description Background. The purpose of this observational case series study was to investigate the role of 25 MHz ultrasound biomicroscopy (UBM) in detecting subluxated lenses and compare it with 50 MHz UBM. Methods. 45 patients (49 eyes) with suspected subluxation of the lens and 20 normal volunteers (40 eyes) were included. Different cross-sectional images of the lens position were captured in axial and longitudinal scanning modes using 25 and 50 MHz UBM. The main outcome measurements included the linear distance between the lens equator and ciliary process, the difference value (D-value) between the same cross section of the above bilateral linear distance in the normal and the subluxated subjects, the diagnostic accuracy, and the testing times obtained with 25 and 50 MHz UBM. Results. The position of the lens on axial sections could be clearly shown by using 25 MHz UBM. The D-value of the subluxated eyes was 1-2 mm longer than that of the normal ones. There was a statistically significant difference between 25 and 50 MHz UBM in showing subluxation of the lens, the testing time was significantly faster (2.0 min versus 7.5 min), and the diagnostic accuracy was much higher (98.0% versus 71.4%) with 25 versus 50 MHz UBM. Fifteen eyes with slightly subluxated lens were detected by 25 MHz UBM, and only one eye with slight lens subluxation was detected by 50 MHz UBM. Conclusions. The results indicated that 25 MHz UBM has a greater diagnostic value than 50 MHz UBM in verifying the status of the lens subluxation and can provide reliable and quantitative imaging evidence for clinical use. This trial is registered with ChiCTR–DOD –15007603.
format Article
id doaj-art-a85fb275266a497d8967a70775a6694b
institution Kabale University
issn 2090-004X
2090-0058
language English
publishDate 2018-01-01
publisher Wiley
record_format Article
series Journal of Ophthalmology
spelling doaj-art-a85fb275266a497d8967a70775a6694b2025-02-03T05:43:41ZengWileyJournal of Ophthalmology2090-004X2090-00582018-01-01201810.1155/2018/37602803760280Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated LensesMingyu Shi0Liwei Ma1Jinsong Zhang2Qichang Yan3Department of Ophthalmology, The Fourth Affiliated Hospital of China Medical University, Eye Hospital of China Medical University, The Key Laboratory of Lens in Liaoning Province, Shenyang, 110005 Liaoning, ChinaDepartment of Ophthalmology, The Fourth Affiliated Hospital of China Medical University, Eye Hospital of China Medical University, The Key Laboratory of Lens in Liaoning Province, Shenyang, 110005 Liaoning, ChinaDepartment of Ophthalmology, The Fourth Affiliated Hospital of China Medical University, Eye Hospital of China Medical University, The Key Laboratory of Lens in Liaoning Province, Shenyang, 110005 Liaoning, ChinaDepartment of Ophthalmology, The Fourth Affiliated Hospital of China Medical University, Eye Hospital of China Medical University, The Key Laboratory of Lens in Liaoning Province, Shenyang, 110005 Liaoning, ChinaBackground. The purpose of this observational case series study was to investigate the role of 25 MHz ultrasound biomicroscopy (UBM) in detecting subluxated lenses and compare it with 50 MHz UBM. Methods. 45 patients (49 eyes) with suspected subluxation of the lens and 20 normal volunteers (40 eyes) were included. Different cross-sectional images of the lens position were captured in axial and longitudinal scanning modes using 25 and 50 MHz UBM. The main outcome measurements included the linear distance between the lens equator and ciliary process, the difference value (D-value) between the same cross section of the above bilateral linear distance in the normal and the subluxated subjects, the diagnostic accuracy, and the testing times obtained with 25 and 50 MHz UBM. Results. The position of the lens on axial sections could be clearly shown by using 25 MHz UBM. The D-value of the subluxated eyes was 1-2 mm longer than that of the normal ones. There was a statistically significant difference between 25 and 50 MHz UBM in showing subluxation of the lens, the testing time was significantly faster (2.0 min versus 7.5 min), and the diagnostic accuracy was much higher (98.0% versus 71.4%) with 25 versus 50 MHz UBM. Fifteen eyes with slightly subluxated lens were detected by 25 MHz UBM, and only one eye with slight lens subluxation was detected by 50 MHz UBM. Conclusions. The results indicated that 25 MHz UBM has a greater diagnostic value than 50 MHz UBM in verifying the status of the lens subluxation and can provide reliable and quantitative imaging evidence for clinical use. This trial is registered with ChiCTR–DOD –15007603.http://dx.doi.org/10.1155/2018/3760280
spellingShingle Mingyu Shi
Liwei Ma
Jinsong Zhang
Qichang Yan
Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
Journal of Ophthalmology
title Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_full Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_fullStr Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_full_unstemmed Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_short Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_sort role of 25 mhz ultrasound biomicroscopy in the detection of subluxated lenses
url http://dx.doi.org/10.1155/2018/3760280
work_keys_str_mv AT mingyushi roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses
AT liweima roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses
AT jinsongzhang roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses
AT qichangyan roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses