Development of Non-Invasive Continuous Blood Pressure Measurement System Based on Multiple Feature Parameters

To address the issue of the difficulty in implementing non-invasive continuous blood pressure measurement technology in China, this study developed a high-performance synchronous electrocardiogram (ECG) and photoplethysmography (PPG) signal acquisition system. A PC-based human-computer interaction s...

Full description

Saved in:
Bibliographic Details
Main Authors: Mingyue LI, Zhiying YUAN, Ruowei LI, Jieying SHAN, Jilun YE, Xu ZHANG, Hui YU, Xiaoyu WU
Format: Article
Language:zho
Published: Editorial Office of Chinese Journal of Medical Instrumentation 2024-11-01
Series:Zhongguo yiliao qixie zazhi
Subjects:
Online Access:https://zgylqxzz.xml-journal.net/article/doi/10.12455/j.issn.1671-7104.230417
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:To address the issue of the difficulty in implementing non-invasive continuous blood pressure measurement technology in China, this study developed a high-performance synchronous electrocardiogram (ECG) and photoplethysmography (PPG) signal acquisition system. A PC-based human-computer interaction software platform was constructed, and continuous blood pressure measurement-related algorithms were integrated. Multiple feature parameters such as pulse wave transit time based on synchronous ECG and PPG signals were extracted, enabling non-invasive continuous blood pressure measurement. To verify the measurement accuracy of the system, tests and comparative verifications were carried out. The results demonstrated that the system could meet the requirements of relevant standards and have good application value.
ISSN:1671-7104