Posterior Dislocation of the Hinge-Post Extension in a Rotating Hinge Total Knee Prosthesis
The rotating hinge knee prosthesis is a popular intervention in patients lacking stability with highly constrained total knee arthroplasty. Despite improvements in design, nonmechanical and mechanical complications continue to be a problem. Dislocation of the hinge has been widely described, mainly...
Saved in:
Main Authors: | , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2013-01-01
|
Series: | Case Reports in Orthopedics |
Online Access: | http://dx.doi.org/10.1155/2013/756538 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The rotating hinge knee prosthesis is a popular intervention in patients lacking stability with highly constrained total knee arthroplasty. Despite improvements in design, nonmechanical and mechanical complications continue to be a problem. Dislocation of the hinge has been widely described, mainly due to the component fracture. Few reports describe isolated dislocation of the rotating stem. We report a case of isolated disengagement of the rotating hinge mechanism, due to severe flexion gap imbalance, leading to subsequent posterior dislocation of the hinge and anterior knee dislocation, in a patient with a history of multiple total knee arthroplasty revisions. This case suggests the importance of the soft tissue balancing, the adequate patellar tracking, and use of a long cylindrical, minimally tapered rotating stem in hinge arthroplasty to minimize hinge dislocation. |
---|---|
ISSN: | 2090-6749 2090-6757 |