Morphological Characterization of the Polyflux 210H Hemodialysis Filter Pores

Background. Morphological characterization of hemodialysis membranes is necessary to improve pore design. Aim. To delineate membrane pore structure of a high flux filter, Polyflux 210H. Methods. We used a Joel JSM-6010LV scanning electron microscope (SEM) and a SU6600 Hitachi field emission scanning...

Full description

Saved in:
Bibliographic Details
Main Authors: A. Hedayat, J. Szpunar, N. A. P. Kiran Kumar, R. Peace, H. Elmoselhi, A. Shoker
Format: Article
Language:English
Published: Wiley 2012-01-01
Series:International Journal of Nephrology
Online Access:http://dx.doi.org/10.1155/2012/304135
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background. Morphological characterization of hemodialysis membranes is necessary to improve pore design. Aim. To delineate membrane pore structure of a high flux filter, Polyflux 210H. Methods. We used a Joel JSM-6010LV scanning electron microscope (SEM) and a SU6600 Hitachi field emission scanning electron microscope (FESEM) to characterize the pore and fiber morphology. The maximal diameters of selected uremic toxins were calculated using the macromolecular modeling Crystallographic Object-Oriented Toolkit (COOT) software. Results. The mean pore densities on the outermost and innermost surfaces of the membrane were 36.81% and 5.45%, respectively. The membrane exhibited a tortuous structure with poor connection between the inner and outer pores. The aperture’s width in the inner surface ranged between 34 and 45 nm, which is 8.76–11.60 times larger than the estimated maximum diameter of β2-microglobulin (3.88 nm). Conclusion. The results suggest that the diameter size of inner pore apertures is not a limiting factor to middle molecules clearance, the extremely diminished density is. Increasing inner pore density and improving channel structure are strategies to improve clearance of middle molecules.
ISSN:2090-214X