Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty
We describe a case of aseptic failure with profound femoral stem trunnion wear and femoral head dissociation nine years after initial primary total hip arthroplasty (THA) with the Stryker Accolade total hip arthroplasty system. Current guidelines for postoperative care and follow-up after THAs as po...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2018-01-01
|
Series: | Case Reports in Orthopedics |
Online Access: | http://dx.doi.org/10.1155/2018/1534572 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832554225944494080 |
---|---|
author | Saif Shamshoon Patrick Thornley Justin de Beer |
author_facet | Saif Shamshoon Patrick Thornley Justin de Beer |
author_sort | Saif Shamshoon |
collection | DOAJ |
description | We describe a case of aseptic failure with profound femoral stem trunnion wear and femoral head dissociation nine years after initial primary total hip arthroplasty (THA) with the Stryker Accolade total hip arthroplasty system. Current guidelines for postoperative care and follow-up after THAs as potential intervention points for early detection of prosthetic joint failure are also reviewed. |
format | Article |
id | doaj-art-7c4196a4ab644b359c66629902c2b43c |
institution | Kabale University |
issn | 2090-6749 2090-6757 |
language | English |
publishDate | 2018-01-01 |
publisher | Wiley |
record_format | Article |
series | Case Reports in Orthopedics |
spelling | doaj-art-7c4196a4ab644b359c66629902c2b43c2025-02-03T05:52:02ZengWileyCase Reports in Orthopedics2090-67492090-67572018-01-01201810.1155/2018/15345721534572Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip ArthroplastySaif Shamshoon0Patrick Thornley1Justin de Beer2Faculty of Health Sciences, Department of Surgery, McMaster University, Hamilton, Ontario, CanadaDivision of Orthopaedic Surgery, McMaster University, Hamilton, Ontario, CanadaDivision of Orthopaedic Surgery, McMaster University, Hamilton, Ontario, CanadaWe describe a case of aseptic failure with profound femoral stem trunnion wear and femoral head dissociation nine years after initial primary total hip arthroplasty (THA) with the Stryker Accolade total hip arthroplasty system. Current guidelines for postoperative care and follow-up after THAs as potential intervention points for early detection of prosthetic joint failure are also reviewed.http://dx.doi.org/10.1155/2018/1534572 |
spellingShingle | Saif Shamshoon Patrick Thornley Justin de Beer Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty Case Reports in Orthopedics |
title | Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty |
title_full | Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty |
title_fullStr | Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty |
title_full_unstemmed | Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty |
title_short | Profound Trunnion Wear Resulting in Femoral Head-Neck Dissociation in Total Hip Arthroplasty |
title_sort | profound trunnion wear resulting in femoral head neck dissociation in total hip arthroplasty |
url | http://dx.doi.org/10.1155/2018/1534572 |
work_keys_str_mv | AT saifshamshoon profoundtrunnionwearresultinginfemoralheadneckdissociationintotalhiparthroplasty AT patrickthornley profoundtrunnionwearresultinginfemoralheadneckdissociationintotalhiparthroplasty AT justindebeer profoundtrunnionwearresultinginfemoralheadneckdissociationintotalhiparthroplasty |