Conventional Rapid Latex Agglutination in Estimation of von Willebrand Factor: Method Revisited and Potential Clinical Applications

Measurement of von Willebrand factor antigen (VWF : Ag) levels is usually performed in a specialised laboratory which limits its application in routine clinical practice. So far, no commercial rapid test kit is available for VWF : Ag estimation. This paper discusses the technical aspect of latex agg...

Full description

Saved in:
Bibliographic Details
Main Authors: Marianor Mahat, Wan Zaidah Abdullah, Che Maraina Che Hussin
Format: Article
Language:English
Published: Wiley 2014-01-01
Series:Journal of Immunology Research
Online Access:http://dx.doi.org/10.1155/2014/850810
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Measurement of von Willebrand factor antigen (VWF : Ag) levels is usually performed in a specialised laboratory which limits its application in routine clinical practice. So far, no commercial rapid test kit is available for VWF : Ag estimation. This paper discusses the technical aspect of latex agglutination method which was established to suit the purpose of estimating von Willebrand factor (VWF) levels in the plasma sample. The latex agglutination test can be performed qualitatively and semiquantitatively. Reproducibility, stability, linearity, limit of detection, interference, and method comparison studies were conducted to evaluate the performance of this test. Semiquantitative latex agglutination test was strongly correlated with the reference immunoturbidimetric assay (Spearman’s rho = 0.946, P<0.001, n=132). A substantial agreement (κ=0.77) was found between qualitative latex agglutination test and the reference assay. Using the scoring system for the rapid latex test, no agglutination is with 0% VWF : Ag (control negative), 1+ reaction is equivalent to <20% VWF : Ag, and 4+ reaction indicates >150% VWF : Ag (when comparing with immunoturbidimetric assay). The findings from evaluation studies suggest that latex agglutination method is suitable to be used as a rapid test kit for the estimation of VWF : Ag levels in various clinical conditions associated with high levels and low levels of VWF : Ag.
ISSN:2314-8861
2314-7156