Nontraumatic Fracture of the Femoral Condylar Prosthesis in a Total Knee Arthroplasty Leading to Mechanical Failure

This paper reports a case of fatigue fracture of the femoral component in a cruciate-retaining cemented total knee arthroplasty (TKA). A 64-year-old man had undergone a primary TKA for osteoarthritis 10 years previously at another institution using the PFC-Sigma prosthesis. The patient recovered ful...

Full description

Saved in:
Bibliographic Details
Main Authors: Girish N. Swamy, Conal Quah, Elmunzar Bagouri, Nitin P. Badhe
Format: Article
Language:English
Published: Wiley 2014-01-01
Series:Case Reports in Orthopedics
Online Access:http://dx.doi.org/10.1155/2014/896348
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This paper reports a case of fatigue fracture of the femoral component in a cruciate-retaining cemented total knee arthroplasty (TKA). A 64-year-old man had undergone a primary TKA for osteoarthritis 10 years previously at another institution using the PFC-Sigma prosthesis. The patient recovered fully and was back to his regular activities. He presented with a history of sudden onset pain and locking of the left knee since the preceding three months. There was no history of trauma, and the patient was mobilizing with difficulty using crutches. Radiographs revealed fracture of the posterior condyle of the femoral prosthesis. Revision surgery was performed as an elective procedure revealing the broken prosthesis. The TC3RP-PFC revision prosthesis was used with a medial parapatellar approach. The patient recovered fully without any squeal. Mechanical failure of the knee arthroplasty prosthesis is rare, and nontraumatic fracture of the femoral metallic component has not been reported before.
ISSN:2090-6749