The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy Needles
Background and Objectives. Recently, a 21G Menghini-type needle for EUS-guided fine-needle aspiration biopsy (EUS-FNAB) has been developed. The stylet of the EUS Sonopsy CY™ remains inside the needle during aspiration. Therefore, it is expected to obtain higher-quality histological core specimens wi...
Saved in:
Main Authors: | , , , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2019-01-01
|
Series: | Gastroenterology Research and Practice |
Online Access: | http://dx.doi.org/10.1155/2019/5810653 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832564330793533440 |
---|---|
author | Ryo Igarashi Atsushi Irisawa Manoop S. Bhutani Irina M. Cazacu Goro Shibukawa Ai Sato Akane Yamabe Takumi Maki Yoshitsugu Yoshida Shogo Yamamoto Tsunehiko Ikeda Hiroshi Hojo |
author_facet | Ryo Igarashi Atsushi Irisawa Manoop S. Bhutani Irina M. Cazacu Goro Shibukawa Ai Sato Akane Yamabe Takumi Maki Yoshitsugu Yoshida Shogo Yamamoto Tsunehiko Ikeda Hiroshi Hojo |
author_sort | Ryo Igarashi |
collection | DOAJ |
description | Background and Objectives. Recently, a 21G Menghini-type needle for EUS-guided fine-needle aspiration biopsy (EUS-FNAB) has been developed. The stylet of the EUS Sonopsy CY™ remains inside the needle during aspiration. Therefore, it is expected to obtain higher-quality histological core specimens without crushing the material or blood contamination. The aim of this study is to evaluate the feasibility and diagnostic accuracy of EUS-FNAB of solid pancreatic masses with this new biopsy needle. Methods. A total of 30 patients with solid pancreatic masses who underwent EUS-FNAB with two different types of needles, EUS Sonopsy™ and ProCore™, were included in a prospective, randomized, controlled, crossover study. All the pancreatic masses were punctured with the two needles and were randomized regarding the order of the needle to be used. The primary outcome was to compare the diagnostic accuracy and the rates of tissue acquisition of the two needles. Results. The tissue acquisition rate was not significantly different between the EUS Sonopsy CY™ needle and the ProCore™ needle (78.6% vs. 82.1%, P=1.00). The histological diagnostic accuracy was also similar between the two needles (73% vs. 80%, P=.63). There was also no difference regarding the accuracy of cytology alone and the combination of both histological and cytological assessments between the EUS Sonopsy CY™ needle and the ProCore™ needle (90% vs. 87%, P=1.00 and 90% vs. 90%, P=1.00, respectively). Conclusions. EUS Sonopsy CY™ is a reliable device for EUS-FNAB of solid pancreatic masses. |
format | Article |
id | doaj-art-63222428e5244716a36b72dabe97259a |
institution | Kabale University |
issn | 1687-6121 1687-630X |
language | English |
publishDate | 2019-01-01 |
publisher | Wiley |
record_format | Article |
series | Gastroenterology Research and Practice |
spelling | doaj-art-63222428e5244716a36b72dabe97259a2025-02-03T01:11:13ZengWileyGastroenterology Research and Practice1687-61211687-630X2019-01-01201910.1155/2019/58106535810653The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy NeedlesRyo Igarashi0Atsushi Irisawa1Manoop S. Bhutani2Irina M. Cazacu3Goro Shibukawa4Ai Sato5Akane Yamabe6Takumi Maki7Yoshitsugu Yoshida8Shogo Yamamoto9Tsunehiko Ikeda10Hiroshi Hojo11Department of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Hepatology and Nutrition, Unit 1466, MD Anderson Cancer Center, University of Texas, Houston, Texas, USADepartment of Gastroenterology, Hepatology and Nutrition, Unit 1466, MD Anderson Cancer Center, University of Texas, Houston, Texas, USADepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Gastroenterology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanDepartment of Pathology, Aizu Medical Center, Fukushima Medical University, Fukushima, JapanBackground and Objectives. Recently, a 21G Menghini-type needle for EUS-guided fine-needle aspiration biopsy (EUS-FNAB) has been developed. The stylet of the EUS Sonopsy CY™ remains inside the needle during aspiration. Therefore, it is expected to obtain higher-quality histological core specimens without crushing the material or blood contamination. The aim of this study is to evaluate the feasibility and diagnostic accuracy of EUS-FNAB of solid pancreatic masses with this new biopsy needle. Methods. A total of 30 patients with solid pancreatic masses who underwent EUS-FNAB with two different types of needles, EUS Sonopsy™ and ProCore™, were included in a prospective, randomized, controlled, crossover study. All the pancreatic masses were punctured with the two needles and were randomized regarding the order of the needle to be used. The primary outcome was to compare the diagnostic accuracy and the rates of tissue acquisition of the two needles. Results. The tissue acquisition rate was not significantly different between the EUS Sonopsy CY™ needle and the ProCore™ needle (78.6% vs. 82.1%, P=1.00). The histological diagnostic accuracy was also similar between the two needles (73% vs. 80%, P=.63). There was also no difference regarding the accuracy of cytology alone and the combination of both histological and cytological assessments between the EUS Sonopsy CY™ needle and the ProCore™ needle (90% vs. 87%, P=1.00 and 90% vs. 90%, P=1.00, respectively). Conclusions. EUS Sonopsy CY™ is a reliable device for EUS-FNAB of solid pancreatic masses.http://dx.doi.org/10.1155/2019/5810653 |
spellingShingle | Ryo Igarashi Atsushi Irisawa Manoop S. Bhutani Irina M. Cazacu Goro Shibukawa Ai Sato Akane Yamabe Takumi Maki Yoshitsugu Yoshida Shogo Yamamoto Tsunehiko Ikeda Hiroshi Hojo The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy Needles Gastroenterology Research and Practice |
title | The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy Needles |
title_full | The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy Needles |
title_fullStr | The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy Needles |
title_full_unstemmed | The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy Needles |
title_short | The Feasibility and Histological Diagnostic Accuracy of Novel Menghini Needle (EUS Sonopsy CY™) for Endoscopic Ultrasound-Guided Fine-Needle Aspiration Biopsy of Solid Pancreatic Masses: A Prospective Crossover Study Comparing Standard Biopsy Needles |
title_sort | feasibility and histological diagnostic accuracy of novel menghini needle eus sonopsy cy™ for endoscopic ultrasound guided fine needle aspiration biopsy of solid pancreatic masses a prospective crossover study comparing standard biopsy needles |
url | http://dx.doi.org/10.1155/2019/5810653 |
work_keys_str_mv | AT ryoigarashi thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT atsushiirisawa thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT manoopsbhutani thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT irinamcazacu thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT goroshibukawa thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT aisato thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT akaneyamabe thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT takumimaki thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT yoshitsuguyoshida thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT shogoyamamoto thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT tsunehikoikeda thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT hiroshihojo thefeasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT ryoigarashi feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT atsushiirisawa feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT manoopsbhutani feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT irinamcazacu feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT goroshibukawa feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT aisato feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT akaneyamabe feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT takumimaki feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT yoshitsuguyoshida feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT shogoyamamoto feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT tsunehikoikeda feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles AT hiroshihojo feasibilityandhistologicaldiagnosticaccuracyofnovelmenghinineedleeussonopsycyforendoscopicultrasoundguidedfineneedleaspirationbiopsyofsolidpancreaticmassesaprospectivecrossoverstudycomparingstandardbiopsyneedles |