Dosimetry and Comparison between Different CT Protocols (Low Dose, Ultralow Dose, and Conventional CT) for Lung Nodules’ Detection in a Phantom

Background. The effects of dose reduction in lung nodule detection need better understanding. Purpose. To compare the detection rate of simulated lung nodules in a chest phantom using different computed tomography protocols, low dose (LD), ultralow dose (ULD), and conventional (CCT), and to quantify...

Full description

Saved in:
Bibliographic Details
Main Authors: Cleverson Alex Leitão, Gabriel Lucca de Oliveira Salvador, Priscilla Tazoniero, Danny Warszawiak, Cristian Saievicz, Rosangela Requi Jakubiak, Dante Luiz Escuissato
Format: Article
Language:English
Published: Wiley 2021-01-01
Series:Radiology Research and Practice
Online Access:http://dx.doi.org/10.1155/2021/6667779
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background. The effects of dose reduction in lung nodule detection need better understanding. Purpose. To compare the detection rate of simulated lung nodules in a chest phantom using different computed tomography protocols, low dose (LD), ultralow dose (ULD), and conventional (CCT), and to quantify their respective amount of radiation. Materials and Methods. A chest phantom containing 93 simulated lung nodules was scanned using five different protocols: ULD (80 kVp/30 mA), LD A (120 kVp/20 mA), LD B (100 kVp/30 mA), LD C (120 kVp/30 mA), and CCT (120 kVp/automatic mA). Four chest radiologists analyzed a selected image from each protocol and registered in diagrams the nodules they detected. Kruskal–Wallis and McNemar’s tests were performed to determine the difference in nodule detection. Equivalent doses were estimated by placing thermoluminescent dosimeters on the surface and inside the phantom. Results. There was no significant difference in lung nodules’ detection when comparing ULD and LD protocols (p=0.208 to p=1.000), but there was a significant difference when comparing each one of those against CCT (p<0.001). The detection rate of nodules with CT attenuation values lower than −600 HU was also different when comparing all protocols against CCT (p<0.001 to p=0.007). There was at least moderate agreement between observers in all protocols (κ-value >0.41). Equivalent dose values ranged from 0.5 to 9 mSv. Conclusion. There is no significant difference in simulated lung nodules’ detection when comparing ULD and LD protocols, but both differ from CCT, especially when considering lower-attenuating nodules.
ISSN:2090-1941
2090-195X