Analytical Method Development and Validation for the Quantification of Acetone and Isopropyl Alcohol in the Tartaric Acid Base Pellets of Dipyridamole Modified Release Capsules by Using Headspace Gas Chromatographic Technique

A simple, sensitive, accurate, robust headspace gas chromatographic method was developed for the quantitative determination of acetone and isopropyl alcohol in tartaric acid-based pellets of dipyridamole modified release capsules. The residual solvents acetone and isopropyl alcohol were used in the...

Full description

Saved in:
Bibliographic Details
Main Authors: Sriram Valavala, Nareshvarma Seelam, Subbaiah Tondepu, V. Shanmukha Kumar Jagarlapudi, Vivekanandan Sundarmurthy
Format: Article
Language:English
Published: Wiley 2018-01-01
Series:Journal of Analytical Methods in Chemistry
Online Access:http://dx.doi.org/10.1155/2018/8240932
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A simple, sensitive, accurate, robust headspace gas chromatographic method was developed for the quantitative determination of acetone and isopropyl alcohol in tartaric acid-based pellets of dipyridamole modified release capsules. The residual solvents acetone and isopropyl alcohol were used in the manufacturing process of the tartaric acid-based pellets of dipyridamole modified release capsules by considering the solubility of the dipyridamole and excipients in the different manufacturing stages. The method was developed and optimized by using fused silica DB-624 (30 m × 0.32 mm × 1.8 µm) column with the flame ionization detector. The method validation was carried out with regard to the guidelines for validation of analytical procedures Q2 demanded by the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH). All the validation characteristics were meeting the acceptance criteria. Hence, the developed and validated method can be applied for the intended routine analysis.
ISSN:2090-8865
2090-8873