METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONS

The high-usage measurement equipment for heat of combustion of organic fuels are bomb isoperibol calorimeters with a water thermostat. The stability of work of calorimeters at real conditions is important for maintenance of reliability of measurement results. The article purpose – the analysis of st...

Full description

Saved in:
Bibliographic Details
Main Authors: Yu. V. Maksimuk, Z. A. Antonava, V. S. Krouk, V. V. Fes’ko, V. N. Kursevich, A. F. Syshchanka, I. V. Vasarenko
Format: Article
Language:English
Published: Belarusian National Technical University 2016-06-01
Series:Приборы и методы измерений
Subjects:
Online Access:https://pimi.bntu.by/jour/article/view/246
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1832557352953315328
author Yu. V. Maksimuk
Z. A. Antonava
V. S. Krouk
V. V. Fes’ko
V. N. Kursevich
A. F. Syshchanka
I. V. Vasarenko
author_facet Yu. V. Maksimuk
Z. A. Antonava
V. S. Krouk
V. V. Fes’ko
V. N. Kursevich
A. F. Syshchanka
I. V. Vasarenko
author_sort Yu. V. Maksimuk
collection DOAJ
description The high-usage measurement equipment for heat of combustion of organic fuels are bomb isoperibol calorimeters with a water thermostat. The stability of work of calorimeters at real conditions is important for maintenance of reliability of measurement results. The article purpose – the analysis of stability for parameters of calorimeters to environment changes. In this work influence room temperature (Тк) and heat exchange conditions on metrological characteristics of two models of calorimeters is considered with different degree of thermal protection: V-08МА and BIC 100. For calorimeters V-08МА the increase in a effective heat capacity (W) on 0,1 % by growth of Tк on everyone 5 °С is established. To use value W in all interval laboratory temperatures Tк = 14–28 °С it is necessary to correct W on 2,8 J/°C on everyone 1 °С changes of Tк. Updating W is required, if the correction exceeds error in determination W. For calorimeter BIC 100 it is not revealed dependences W from Tк. BIC 100 have constant-temperature cap, high stability a temperature in thermostat and stabilized heat exchange. It is established that an standard deviation of cooling constant for all calorimeters in direct proportional to standard deviation W.
format Article
id doaj-art-1493acc7e7d5494fbdbf0d8790128df9
institution Kabale University
issn 2220-9506
2414-0473
language English
publishDate 2016-06-01
publisher Belarusian National Technical University
record_format Article
series Приборы и методы измерений
spelling doaj-art-1493acc7e7d5494fbdbf0d8790128df92025-02-03T05:16:53ZengBelarusian National Technical UniversityПриборы и методы измерений2220-95062414-04732016-06-01719510310.21122/2220-9506-2016-7-1-123-125235METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONSYu. V. Maksimuk0Z. A. Antonava1V. S. Krouk2V. V. Fes’ko3V. N. Kursevich4A. F. Syshchanka5I. V. Vasarenko6Research Institute for Physical Chemical Problems of the Belarusian State UniversityResearch Institute for Physical Chemical Problems of the Belarusian State UniversityResearch Institute for Physical Chemical Problems of the Belarusian State UniversityResearch Institute for Physical Chemical Problems of the Belarusian State UniversityResearch Institute for Physical Chemical Problems of the Belarusian State UniversityBMC JSC, MinskBMC JSC, MinskThe high-usage measurement equipment for heat of combustion of organic fuels are bomb isoperibol calorimeters with a water thermostat. The stability of work of calorimeters at real conditions is important for maintenance of reliability of measurement results. The article purpose – the analysis of stability for parameters of calorimeters to environment changes. In this work influence room temperature (Тк) and heat exchange conditions on metrological characteristics of two models of calorimeters is considered with different degree of thermal protection: V-08МА and BIC 100. For calorimeters V-08МА the increase in a effective heat capacity (W) on 0,1 % by growth of Tк on everyone 5 °С is established. To use value W in all interval laboratory temperatures Tк = 14–28 °С it is necessary to correct W on 2,8 J/°C on everyone 1 °С changes of Tк. Updating W is required, if the correction exceeds error in determination W. For calorimeter BIC 100 it is not revealed dependences W from Tк. BIC 100 have constant-temperature cap, high stability a temperature in thermostat and stabilized heat exchange. It is established that an standard deviation of cooling constant for all calorimeters in direct proportional to standard deviation W.https://pimi.bntu.by/jour/article/view/246bomb calorimetereffective heat capacitycooling constantheat leak correction
spellingShingle Yu. V. Maksimuk
Z. A. Antonava
V. S. Krouk
V. V. Fes’ko
V. N. Kursevich
A. F. Syshchanka
I. V. Vasarenko
METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONS
Приборы и методы измерений
bomb calorimeter
effective heat capacity
cooling constant
heat leak correction
title METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONS
title_full METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONS
title_fullStr METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONS
title_full_unstemmed METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONS
title_short METROLOGICAL PERFORMANCES OF BOMB CALORIMETERS AT REAL CONDITIONS
title_sort metrological performances of bomb calorimeters at real conditions
topic bomb calorimeter
effective heat capacity
cooling constant
heat leak correction
url https://pimi.bntu.by/jour/article/view/246
work_keys_str_mv AT yuvmaksimuk metrologicalperformancesofbombcalorimetersatrealconditions
AT zaantonava metrologicalperformancesofbombcalorimetersatrealconditions
AT vskrouk metrologicalperformancesofbombcalorimetersatrealconditions
AT vvfesko metrologicalperformancesofbombcalorimetersatrealconditions
AT vnkursevich metrologicalperformancesofbombcalorimetersatrealconditions
AT afsyshchanka metrologicalperformancesofbombcalorimetersatrealconditions
AT ivvasarenko metrologicalperformancesofbombcalorimetersatrealconditions