Ultrasonic Mammography with Circular Transducer Array

Ultrasonic projection imaging is similar to X-ray radiography. Nowadays, ultrasonic projection methods have been developed in the set-up of multi-element flat arrays with miniature transducers, where one of the array acts as a transmitter and the other one is a receiver. In the paper, a new metho...

Full description

Saved in:
Bibliographic Details
Main Authors: Krzysztof J. OPIELIŃSKI, Piotr PRUCHNICKI, Tadeusz GUDRA
Format: Article
Language:English
Published: Institute of Fundamental Technological Research Polish Academy of Sciences 2014-09-01
Series:Archives of Acoustics
Subjects:
Online Access:https://acoustics.ippt.pan.pl/index.php/aa/article/view/1435
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1850104855458217984
author Krzysztof J. OPIELIŃSKI
Piotr PRUCHNICKI
Tadeusz GUDRA
author_facet Krzysztof J. OPIELIŃSKI
Piotr PRUCHNICKI
Tadeusz GUDRA
author_sort Krzysztof J. OPIELIŃSKI
collection DOAJ
description Ultrasonic projection imaging is similar to X-ray radiography. Nowadays, ultrasonic projection methods have been developed in the set-up of multi-element flat arrays with miniature transducers, where one of the array acts as a transmitter and the other one is a receiver. In the paper, a new method of the projection imaging using a 1024-element circular ultrasonic transducer array was presented. It allows the choice of a projection scanning plane for any angle around a studied object submerged in water. Fast acquisition of measurement data is achieved as a result of parallel switching of opposite transmitting and receiving transducers in the circular array and vertical movement of the array. The algorithm equalizing the length of measurement rays and distances between them was elaborated for the reconstruction of projection images. Projection research results of breast phantom obtained by means of the elaborated measurement set-up and compared with mammography simulations (acquired through overlapping of X-ray tomographic images) show that ultrasonic projection method presented in this paper (so-called ultrasonic mammography) can be applied to the woman’s breast and be used as a diagnosis for an early detection of cancerous lesions. It can, most of all, be used as an alternative or complementary method to standard mammography, which is harmful because of ionizing radiation and invasive because of the mechanical compression of tissue.
format Article
id doaj-art-10552b60b26d4d78a63995666170c4d3
institution DOAJ
issn 0137-5075
2300-262X
language English
publishDate 2014-09-01
publisher Institute of Fundamental Technological Research Polish Academy of Sciences
record_format Article
series Archives of Acoustics
spelling doaj-art-10552b60b26d4d78a63995666170c4d32025-08-20T02:39:15ZengInstitute of Fundamental Technological Research Polish Academy of SciencesArchives of Acoustics0137-50752300-262X2014-09-0139410.2478/aoa-2014-0060Ultrasonic Mammography with Circular Transducer ArrayKrzysztof J. OPIELIŃSKI0Piotr PRUCHNICKI1Tadeusz GUDRA2Chair of Acoustics and Multimedia, Faculty of Electronics, Wrocław University of TechnologyChair of Acoustics and Multimedia, Faculty of Electronics, Wrocław University of TechnologyChair of Acoustics and Multimedia, Faculty of Electronics, Wrocław University of TechnologyUltrasonic projection imaging is similar to X-ray radiography. Nowadays, ultrasonic projection methods have been developed in the set-up of multi-element flat arrays with miniature transducers, where one of the array acts as a transmitter and the other one is a receiver. In the paper, a new method of the projection imaging using a 1024-element circular ultrasonic transducer array was presented. It allows the choice of a projection scanning plane for any angle around a studied object submerged in water. Fast acquisition of measurement data is achieved as a result of parallel switching of opposite transmitting and receiving transducers in the circular array and vertical movement of the array. The algorithm equalizing the length of measurement rays and distances between them was elaborated for the reconstruction of projection images. Projection research results of breast phantom obtained by means of the elaborated measurement set-up and compared with mammography simulations (acquired through overlapping of X-ray tomographic images) show that ultrasonic projection method presented in this paper (so-called ultrasonic mammography) can be applied to the woman’s breast and be used as a diagnosis for an early detection of cancerous lesions. It can, most of all, be used as an alternative or complementary method to standard mammography, which is harmful because of ionizing radiation and invasive because of the mechanical compression of tissue.https://acoustics.ippt.pan.pl/index.php/aa/article/view/1435ultrasonic projection imaging (UPI)ultrasonic mammographycircular transducer arraywoman’s breast tissue.
spellingShingle Krzysztof J. OPIELIŃSKI
Piotr PRUCHNICKI
Tadeusz GUDRA
Ultrasonic Mammography with Circular Transducer Array
Archives of Acoustics
ultrasonic projection imaging (UPI)
ultrasonic mammography
circular transducer array
woman’s breast tissue.
title Ultrasonic Mammography with Circular Transducer Array
title_full Ultrasonic Mammography with Circular Transducer Array
title_fullStr Ultrasonic Mammography with Circular Transducer Array
title_full_unstemmed Ultrasonic Mammography with Circular Transducer Array
title_short Ultrasonic Mammography with Circular Transducer Array
title_sort ultrasonic mammography with circular transducer array
topic ultrasonic projection imaging (UPI)
ultrasonic mammography
circular transducer array
woman’s breast tissue.
url https://acoustics.ippt.pan.pl/index.php/aa/article/view/1435
work_keys_str_mv AT krzysztofjopielinski ultrasonicmammographywithcirculartransducerarray
AT piotrpruchnicki ultrasonicmammographywithcirculartransducerarray
AT tadeuszgudra ultrasonicmammographywithcirculartransducerarray