Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography
Purpose. To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes. Method...
Saved in:
Main Authors: | , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2020-01-01
|
Series: | Journal of Ophthalmology |
Online Access: | http://dx.doi.org/10.1155/2020/9739878 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832551233823440896 |
---|---|
author | Sisi Chen Rongrong Gao Colm McAlinden Junming Ye Yiran Wang Min Chen Jinhai Huang Yong Sun A-Yong Yu |
author_facet | Sisi Chen Rongrong Gao Colm McAlinden Junming Ye Yiran Wang Min Chen Jinhai Huang Yong Sun A-Yong Yu |
author_sort | Sisi Chen |
collection | DOAJ |
description | Purpose. To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes. Methods. Sixty-eight eyes of 68 subjects were included. Three consecutive scans of each subject were obtained using both devices in a random order by one experienced operator. Standard deviation (Sw), coefficient of repeatability (CoR), coefficients of variation (CoV), and intraclass correlation coefficients (ICC) were used to evaluate the intraoperator repeatability. Agreement was assessed using the Bland–Altman plots and 95% limits of agreement (LoA). Results. All measurements of the swept-source OCT (SS-OCT) and time-domain OCT (TD-OCT) showed high repeatability with low CoR (CCT: 2.34 μm and 6.16 μm; AQD: 0.05 mm and 0.09 mm; ACD: 0.06 mm and 0.09 mm), low CoV (CCT: 0.16% and 0.42%; AQD: 0.61% and 0.97%; ACD: 0.53% and 0.83%), and high ICC (>0.98). The mean CCT with SS-OCT was slightly thicker than the results with TD-OCT (difference = 4.55 ± 2.62 μm, P<0.001). There was no statistically significant difference in AQD or ACD measurements between the two devices (0.01 ± 0.05 mm, P=0.111; 0.02 ± 0.05 mm, P=0.022, respectively). The 95% LoA between the SS-OCT and TD-OCT were −0.59 to 9.69 μm for CCT, −0.10 to 0.12 mm for AQD, and −0.09 to 0.12 mm for ACD. Conclusions. High levels of repeatability and agreement were found between the two devices for all three parameters, suggesting interchangeability. SS-OCT demonstrated superior repeatability compared with TD-OCT. |
format | Article |
id | doaj-art-1046c2b22a044539a334210c2c4b1442 |
institution | Kabale University |
issn | 2090-004X 2090-0058 |
language | English |
publishDate | 2020-01-01 |
publisher | Wiley |
record_format | Article |
series | Journal of Ophthalmology |
spelling | doaj-art-1046c2b22a044539a334210c2c4b14422025-02-03T06:04:35ZengWileyJournal of Ophthalmology2090-004X2090-00582020-01-01202010.1155/2020/97398789739878Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence TomographySisi Chen0Rongrong Gao1Colm McAlinden2Junming Ye3Yiran Wang4Min Chen5Jinhai Huang6Yong Sun7A-Yong Yu8School of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaShenzhen Hospital of Integrated Traditional and Western Medicine, Shenzhen, ChinaSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaPurpose. To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes. Methods. Sixty-eight eyes of 68 subjects were included. Three consecutive scans of each subject were obtained using both devices in a random order by one experienced operator. Standard deviation (Sw), coefficient of repeatability (CoR), coefficients of variation (CoV), and intraclass correlation coefficients (ICC) were used to evaluate the intraoperator repeatability. Agreement was assessed using the Bland–Altman plots and 95% limits of agreement (LoA). Results. All measurements of the swept-source OCT (SS-OCT) and time-domain OCT (TD-OCT) showed high repeatability with low CoR (CCT: 2.34 μm and 6.16 μm; AQD: 0.05 mm and 0.09 mm; ACD: 0.06 mm and 0.09 mm), low CoV (CCT: 0.16% and 0.42%; AQD: 0.61% and 0.97%; ACD: 0.53% and 0.83%), and high ICC (>0.98). The mean CCT with SS-OCT was slightly thicker than the results with TD-OCT (difference = 4.55 ± 2.62 μm, P<0.001). There was no statistically significant difference in AQD or ACD measurements between the two devices (0.01 ± 0.05 mm, P=0.111; 0.02 ± 0.05 mm, P=0.022, respectively). The 95% LoA between the SS-OCT and TD-OCT were −0.59 to 9.69 μm for CCT, −0.10 to 0.12 mm for AQD, and −0.09 to 0.12 mm for ACD. Conclusions. High levels of repeatability and agreement were found between the two devices for all three parameters, suggesting interchangeability. SS-OCT demonstrated superior repeatability compared with TD-OCT.http://dx.doi.org/10.1155/2020/9739878 |
spellingShingle | Sisi Chen Rongrong Gao Colm McAlinden Junming Ye Yiran Wang Min Chen Jinhai Huang Yong Sun A-Yong Yu Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography Journal of Ophthalmology |
title | Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography |
title_full | Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography |
title_fullStr | Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography |
title_full_unstemmed | Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography |
title_short | Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography |
title_sort | comparison of anterior ocular biometric measurements using swept source and time domain optical coherence tomography |
url | http://dx.doi.org/10.1155/2020/9739878 |
work_keys_str_mv | AT sisichen comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT rongronggao comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT colmmcalinden comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT junmingye comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT yiranwang comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT minchen comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT jinhaihuang comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT yongsun comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography AT ayongyu comparisonofanteriorocularbiometricmeasurementsusingsweptsourceandtimedomainopticalcoherencetomography |