Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device
Background. The potential of pocket-sized ultrasound devices (PUDs) to improve global healthcare delivery is limited by the lack of a suitable imaging protocol and trained users. Therefore, we investigated the feasibility of performing a brief, evidence-based cardiac limited ultrasound exam (CLUE) t...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2013-01-01
|
Series: | Emergency Medicine International |
Online Access: | http://dx.doi.org/10.1155/2013/627230 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1832556585437626368 |
---|---|
author | Tuan V. Mai David T. Ahn Colin T. Phillips Donna L. Agan Bruce J. Kimura |
author_facet | Tuan V. Mai David T. Ahn Colin T. Phillips Donna L. Agan Bruce J. Kimura |
author_sort | Tuan V. Mai |
collection | DOAJ |
description | Background. The potential of pocket-sized ultrasound devices (PUDs) to improve global healthcare delivery is limited by the lack of a suitable imaging protocol and trained users. Therefore, we investigated the feasibility of performing a brief, evidence-based cardiac limited ultrasound exam (CLUE) through wireless guidance of novice users. Methods. Three trainees applied PUDs on 27 subjects while directed by an off-site cardiologist to obtain a CLUE to screen for LV systolic dysfunction (LVSD), LA enlargement (LAE), ultrasound lung comets (ULC+), and elevated CVP (eCVP). Real-time remote audiovisual guidance and interpretation by the cardiologist were performed using the iPhone 4/iPod (FaceTime, Apple, Inc.) attached to the PUD and transmitted data wirelessly. Accuracy and technical quality of transmitted images were compared to on-site, gold-standard echo thresholds. Results. Novice versus sonographer imaging yielded technically adequate views in 122/135 (90%) versus 130/135 (96%) (). CLUE’s combined SN, SP, and ACC were 0.67, 0.96, and 0.90. Technical adequacy (%) and accuracy for each abnormality () were LVSD (85%, 0.93, ), LAE (89%, 0.74, ), ULC+ (100%, 0.94, ), and eCVP (78%, 0.91, ). Conclusion. A novice can perform the CLUE using PUD when wirelessly guided by an expert. This method could facilitate PUD use for off-site bedside medical decision making and triaging of patients. |
format | Article |
id | doaj-art-013fad3844a44f0e9c77c86b8f6d5966 |
institution | Kabale University |
issn | 2090-2840 2090-2859 |
language | English |
publishDate | 2013-01-01 |
publisher | Wiley |
record_format | Article |
series | Emergency Medicine International |
spelling | doaj-art-013fad3844a44f0e9c77c86b8f6d59662025-02-03T05:44:56ZengWileyEmergency Medicine International2090-28402090-28592013-01-01201310.1155/2013/627230627230Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound DeviceTuan V. Mai0David T. Ahn1Colin T. Phillips2Donna L. Agan3Bruce J. Kimura4Department of Graduate Medical Education, Scripps Mercy Hospital, 4077 Fifth Avenue, MER-35, San Diego, CA 92103, USADepartment of Graduate Medical Education, Scripps Mercy Hospital, 4077 Fifth Avenue, MER-35, San Diego, CA 92103, USADepartment of Graduate Medical Education, Scripps Mercy Hospital, 4077 Fifth Avenue, MER-35, San Diego, CA 92103, USADepartment of Graduate Medical Education, Scripps Mercy Hospital, 4077 Fifth Avenue, MER-35, San Diego, CA 92103, USADepartment of Graduate Medical Education, Scripps Mercy Hospital, 4077 Fifth Avenue, MER-35, San Diego, CA 92103, USABackground. The potential of pocket-sized ultrasound devices (PUDs) to improve global healthcare delivery is limited by the lack of a suitable imaging protocol and trained users. Therefore, we investigated the feasibility of performing a brief, evidence-based cardiac limited ultrasound exam (CLUE) through wireless guidance of novice users. Methods. Three trainees applied PUDs on 27 subjects while directed by an off-site cardiologist to obtain a CLUE to screen for LV systolic dysfunction (LVSD), LA enlargement (LAE), ultrasound lung comets (ULC+), and elevated CVP (eCVP). Real-time remote audiovisual guidance and interpretation by the cardiologist were performed using the iPhone 4/iPod (FaceTime, Apple, Inc.) attached to the PUD and transmitted data wirelessly. Accuracy and technical quality of transmitted images were compared to on-site, gold-standard echo thresholds. Results. Novice versus sonographer imaging yielded technically adequate views in 122/135 (90%) versus 130/135 (96%) (). CLUE’s combined SN, SP, and ACC were 0.67, 0.96, and 0.90. Technical adequacy (%) and accuracy for each abnormality () were LVSD (85%, 0.93, ), LAE (89%, 0.74, ), ULC+ (100%, 0.94, ), and eCVP (78%, 0.91, ). Conclusion. A novice can perform the CLUE using PUD when wirelessly guided by an expert. This method could facilitate PUD use for off-site bedside medical decision making and triaging of patients.http://dx.doi.org/10.1155/2013/627230 |
spellingShingle | Tuan V. Mai David T. Ahn Colin T. Phillips Donna L. Agan Bruce J. Kimura Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device Emergency Medicine International |
title | Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device |
title_full | Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device |
title_fullStr | Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device |
title_full_unstemmed | Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device |
title_short | Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device |
title_sort | feasibility of remote real time guidance of a cardiac examination performed by novices using a pocket sized ultrasound device |
url | http://dx.doi.org/10.1155/2013/627230 |
work_keys_str_mv | AT tuanvmai feasibilityofremoterealtimeguidanceofacardiacexaminationperformedbynovicesusingapocketsizedultrasounddevice AT davidtahn feasibilityofremoterealtimeguidanceofacardiacexaminationperformedbynovicesusingapocketsizedultrasounddevice AT colintphillips feasibilityofremoterealtimeguidanceofacardiacexaminationperformedbynovicesusingapocketsizedultrasounddevice AT donnalagan feasibilityofremoterealtimeguidanceofacardiacexaminationperformedbynovicesusingapocketsizedultrasounddevice AT brucejkimura feasibilityofremoterealtimeguidanceofacardiacexaminationperformedbynovicesusingapocketsizedultrasounddevice |